Connecting the Cables
A-6 Setting Up Your Cardiograph
Connecting the Cables
1. Plug the power cord into the wall outlet.
2. Connect the power cord to the cardiograph as shown in Figure A-4.
Figure A-4: Connecting the Power Cord.
A. Power Cord
B. Voltage Select Switch
WARNI NG If you must use an ungrounded plug adapter to plug the power cord into the wall
outlet, you must also use a ground strap to connect the equipotential connector at
the rear of the cardiograph to the power source ground.
Figure A-1 shows the location of the equipotential connector. Refer to the Patient
and Operational Safety Notes in Getting Acquainted for more information about
using the equipotential connector.