Agilent 3070 / 79000
DUT Power Supply Installation
October 2003
Zobrazit stránku 0
1 2 3 4 5 6 ... 81 82

Shrnutí obsahu

Strany 1 - DUT Power Supply Installation

Agilent 3070 / 79000 DUT Power Supply InstallationOctober 2003

Strany 2

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 7IntroductionFigure 1 Illustrations of cable ends. Top left: DUT

Strany 3 - Table of Contents

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 8Setting the DUT Power Supply Line VoltageSetting the DUT Power S

Strany 4

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 9Setting the DUT Power Supply Line VoltageThe Testhead product nu

Strany 5

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 10Setting the DUT Power Supply Line VoltageTable 3 DUT power supp

Strany 6 - Preparation

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 11Setting the DUT Power Supply Line VoltageSingle-Phase Earthed 0

Strany 7

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 12Setting the DUT Power Supply Line VoltageNOTEOption 05A is not

Strany 8

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 13Setting the DUT Power Supply Line Voltage6621A and 6624A Line V

Strany 9 - 1. Series I

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 14Setting the DUT Power Supply Line VoltageThe procedure for chan

Strany 10 - Introduction

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 15Setting the DUT Power Supply Line VoltageFigure 3 6624A and 662

Strany 11 - Power Supply Line

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 16Setting the DUT Power Supply Line Voltage5 Install the correct

Strany 12

© Agilent Technologies 2000, 2002, 2003 Agilent 3070 / 79000 DUT Power Supply Installation 1ContentsDUT Power Supply InstallationE4000-90021 Rev. J

Strany 13

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 17Setting the DUT Power Supply Line VoltageFigure 4 6634A line vo

Strany 14

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 18Setting the DUT Power Supply Line Voltage3 Note that each swit

Strany 15

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 19Setting the DUT Power Supply Line VoltageFigure 5 6634B line vo

Strany 16 - Line input power module

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 20Setting the DUT Power Supply Line Voltage6642A Line Voltage Set

Strany 17

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 21Setting the DUT Power Supply Line VoltageWARNING✸Make sure the

Strany 18 - Line input

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 22Installing DUT Power SuppliesInstalling DUT Power SuppliesThis

Strany 19 - 6634A Line Voltage Settings

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 23Installing DUT Power SuppliesFigure 7 Radiation shields for the

Strany 20 - 6632ps.wpg

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 24Installing DUT Power SuppliesInstalling the DUT Power Supply in

Strany 21 - 6634B Line Voltage Settings

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 25Installing DUT Power SuppliesFigure 8 Locating the DUT power su

Strany 22

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 26Installing DUT Power Supplies6 Using a #2 Pozidriv screwdriver

Strany 23 - 6642pow.wpg

© Agilent Technologies 2000, 2002, 2003 Agilent 3070 / 79000 DUT Power Supply Installation 2Table of ContentsVerifying Operation...

Strany 24

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 27Installing DUT Power SuppliesFigure 9 Front view of the testhea

Strany 25 - Power Supplies

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 28Installing DUT Power Supplies8 Slide the power supply into the

Strany 26 - Installing DUT Power Supplies

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 29Installing DUT Power SuppliesFigure 10 Installing the DUT power

Strany 27

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 30Installing DUT Power SuppliesInstalling the DUT Power Supply in

Strany 28 - Support Rails

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 31Installing DUT Power SuppliesFigure 11 327X DUT power supply lo

Strany 29

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 32Installing DUT Power Supplies3 Determine whether the radiation

Strany 30

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 33Installing DUT Power SuppliesFigure 12 Location of the DUT powe

Strany 31

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 34Installing DUT Power SuppliesFigure 13). Pan-head screws are in

Strany 32

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 35Installing DUT Power SuppliesInstalling the DUT Power Supplies

Strany 33

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 36Installing DUT Power SuppliesFigure 14 79000 DUT power supply l

Strany 34

© Agilent Technologies 2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 1 DUT Power Supply InstallationE4000-90021 Rev. J 10/2003In

Strany 35

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 37Installing DUT Power SuppliesMiddle 6634A/B DUT Supply Installa

Strany 36 - 327Xshield.wpg

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 38Installing DUT Power Supplies5 Install the 6634A/B in the midd

Strany 37 - Front of Power Supply

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 39Installing DUT Power SuppliesFigure 17 Second 6634A front mount

Strany 38 - Figure 14 on page 36. The

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 40Installing DUT Power SuppliesFigure 18 6624A rear mounting brac

Strany 39

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 41Connecting the DUT Power Supply CablesConnecting the DUT Power

Strany 40

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 42Connecting the DUT Power Supply CablesThe cables used to connec

Strany 41 - Rack Mount Ears

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 43Connecting the DUT Power Supply CablesFigure 19 ASRU card conne

Strany 42 - Cradle Frame

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 44Connecting the DUT Power Supply CablesFigure 20 Left rear view

Strany 43

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 45Connecting the DUT Power Supply Cables

Strany 44 - Power Supply

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 46Connecting the DUT Power Supply CablesNOTEThe pins in the DUT p

Strany 45

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 2IntroductionIntroduction This section contains: Introduction, 2

Strany 46 - ASRU Card

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 47Connecting the DUT Power Supply Cables6624A (44940A and E3785A)

Strany 47 - DUTback.wpg

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 48Connecting the DUT Power Supply CablesFigure 21 6624A output co

Strany 48

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 49Connecting the DUT Power Supply Cables2 If channels 1 through

Strany 49

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 50Connecting the DUT Power Supply CablesFigure 22 Alternate 6624A

Strany 50 - Figure 21

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 51Connecting the DUT Power Supply Cables6621A (44941A) Supply The

Strany 51

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 52Connecting the DUT Power Supply CablesFigure 23 6621A output co

Strany 52 - Figure 22 on page 50. Notice

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 53Connecting the DUT Power Supply Cables2 If you are using ASRU

Strany 53 - 20-pin connector

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 54Connecting the DUT Power Supply CablesFigure 24 Alternate 6621A

Strany 54 - Figure 23 on

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 55Connecting the DUT Power Supply Cables6634A/B (E4034A) Supply T

Strany 55 - DUT_PS4.WPG

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 56Connecting the DUT Power Supply Cables6642A (E3784A) Supply The

Strany 56 - Figure 24 on page 54

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 3Introduction4 Update the system configuration file for the addi

Strany 57

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 57Connecting the DUT Power Supply CablesConnect the cables to the

Strany 58 - Figure 37 on page 75

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 58Connecting the DUT Power Supply CablesFigure 27 Agilent 6642A o

Strany 59 - Channel 6

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 59Connecting the DUT Power Supply CablesConnecting the Overvoltag

Strany 60 - Figure 27

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 60Connecting the DUT Power Supply Cables(03066-61622) has two 2-w

Strany 61 - +V -V GND

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 61Connecting the DUT Power Supply CablesFigure 28 DUT power suppl

Strany 62

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 62Connecting the DUT Power Supply CablesFigure 29 DUT power suppl

Strany 63

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 63Connecting the DUT Power Supply CablesFigure 30 DUT power suppl

Strany 64 - 03066-61622

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 64Connecting the DUT Power Supply CablesFigure 31 DUT power suppl

Strany 65

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 65Connecting the DUT Power Supply CablesFigure 32 DUT power suppl

Strany 66

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 66Connecting the DUT Power Supply CablesFigure 33 DUT power suppl

Strany 67 - One DUT Per Module

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 4IntroductionTools for removing the handler: 2.5-millimeter hex

Strany 68

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 67Connecting the DUT Power Supply CablesFigure 34 DUT power suppl

Strany 69 - 3066-61622

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 68Connecting the DUT Power Supply CablesFigure 35 DUT power suppl

Strany 70

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 69Connecting the DUT Power Supply CablesConnecting the Power and

Strany 71

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 70Connecting the DUT Power Supply CablesFigure 36 Connecting the

Strany 72

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 71Connecting the DUT Power Supply Cables79K-HPIB.WPGInstrument Ra

Strany 73

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 72Connecting the DUT Power Supply CablesProgramming the GPIB Addr

Strany 74 - 79K-HPIB.WPG

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 73Connecting the DUT Power Supply CablesProgramming the GPIB Addr

Strany 75

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 74Connecting the DUT Power Supply CablesConnecting Output, Overvo

Strany 76

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 75Connecting the DUT Power Supply CablesFigure 37 79000 DUT power

Strany 77 - GPIB Cables in the 79000

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 76Completing the InstallationCompleting the InstallationThis sect

Strany 78 - ASRU card

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 5Introduction8 For 327X systems, remove the testhead covers (sav

Strany 79 - Installation

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 77Completing the Installationphysical connection of the supply to

Strany 80

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 78Completing the InstallationVerifying Operation 1 Look at the T

Strany 81 - Replacing the Covers

Printed in SingaporeE4000-90021 Rev. J 10/2003

Strany 82 - E4000-90021 Rev. J 10/2003

© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 6Introduction.Table 1 DUT power supply output cable part numbersS