© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 76
Completing the Installation
Completing the
Installation
This section contains:
6642A - Changing the Programming Mode, 76
Updating the Configuration Files, 76
Verifying Operation, 78
In Case of Difficulty, 78
Replacing the Covers, 78
Installation Billing Information, 78
6642A - Changing the Programming Mode
The 6642A defaults to TMSL/SCPI
1
programming
mode. If the power supply is installed at the factory in a
3070 / 79000 system, the mode will be set correctly;
however exchanged or new 6642A power supplies must
be set to COMP (compatibility) mode. The 3070 / 79000
cannot communicate with the 6642A in TMSL/SCPI
programming mode.
To set the programming mode to COMP mode, install
the 6642A in the system according to the previous
instructions. Then, from a BT-BASIC window, execute
the following commands:
assign @ps<#> to btgetenv$(“AGILENT3070_ROOT”)&“/dev/ps<#>” ! for MS Windows system
assign @ps<#> to “/dev/ps<#>” ! for UNIX system
output @ps<#>; “SYST:LANG COMP”
Updating the Configuration Files
The power supply connections must be defined in the
system and standard config files:
UX:
/var/hp3070/diagnostics/th1/config and
/var/hp3070/standard/config
WN: \Agilent3070\diagnostics\th1\config and
\Agilent3070\standard\config
Use the supplies statement.
supplies <type> <#> asru channel <#>
The type is either hp6624, hp6621, hp6634, or hp6642
depending on which supply was added.
The first number or range of numbers gives the testhead
supply number (1 through 24). The numbering of the
supplies in the
supplies statement is arbitrary. The
software will accept any mapping of supply numbers to
modules. However, you cannot assign a power supply
number more than once in a testhead. If the customer
has more than one system, make the new
supplies
statement compatible with the other system.
The second number or range of numbers gives the
ASRU Card DUT channel. This must match the
1 Test and Measurement System Language/Standard Commands for Programmable Instruments