Agilent Technologies HP 3325B Manuál s instrukcemi Strana 128

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 252
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 127
Chapter 2 Service
About the Agilent 53181A Calibration Menu
2-30 Assembly-Level Service Guide
2
To Change to New the Security Code
1 Press and hold Scale & Offset key, then cycle POWER key.
2 Unsecure for calibration by the using the procedure described in the section
titled “To Unsecure for Calibration.
3 Press Scale & Offset key until CODE: is displayed.
4 Press the appropriate arrow keys to enter the new or customized security code,
then press Enter key.
A scrolling message will appear, indicating what the new code is. For example, the
Counter displays the NEW CODE IS 5 if you entered “5” as the new code.
To View the Calibration Count
1 Press and hold Scale & Offset key, then cycle POWER key.
2 Press Scale & Offset key until CAL COUNT? is displayed.
3 Press any one of the entry keys (arrow, +/-, Enter) to have a scrolling message
appear, indicating what the calibration count is.
Your Counter was calibrated before it left the factory. When you receive your
Counter, read the calibration count to determine its initial value.
To Get Help with the Calibration Menu
1 Press and hold Scale & Offset key, then cycle POWER key.
2 Press Scale & Offset key until HELP: CAL? is displayed.
3 Press any one of the arrow keys until your help choice (that is, HELP: CAL?,
HELP: SECURE?, HELP: CODE?) is displayed.
4 Press Enter key to display the help information.
Zobrazit stránku 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 251 252

Komentáře k této Příručce

Žádné komentáře