Agilent Technologies HP 3325B Manuál s instrukcemi Strana 45

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 252
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 44
Chapter 1 Performance Tests
Test 2: Trigger Level (Agilent 53131A/132A Only)
Assembly-Level Service Guide 1-21
1
7 In the Counter’s Channel 1 Trigger/Sensitivity menu, change the SLOPE to
NEG.
8 Set the Agilent 3325B DC OFFSET to +60 mV.
9 On the Counter, press Run key.
10 Now, decrement the offset on the Agilent 3325B by 1 mV until the Counters
Gate annunciator flashes and continue decrementing until the Counter displays
a stable reading of approximately 1 MHz.
Observe the offset value on the display of the Agilent 3325B.
Record the value__________ mV.
11 Add the lower peak voltage (40mV) of the 80 mVp-p signal to the offset value in
step 10 (for example, 21 mV 40 mV = 19 mv.)
This result is the lower hysteresis level.
Record the lower hysteresis level __________ mV.
12 Now, add the upper hysteresis value (recorded in step 6) and lower hysteresis
value (recorded in step 11).
This is the trigger level; it should be 0.0 V ±15 mV.
13 Record the trigger level value in the appropriate place in the Performance Test
Record (Test 2, Line 1).
14 Repeat steps 1 through 13 for Channel 2.
15 Record the trigger level value in the appropriate place in the Performance Test
Record (Test 2, Line 2).
16 Disconnect the test setup.
Zobrazit stránku 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 251 252

Komentáře k této Příručce

Žádné komentáře