Agilent Technologies HP 3325B Manuál s instrukcemi Strana 212

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 252
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 211
Chapter 6 Backdating
The Agilent 53131A Calibration Procedures
6-10 Assembly-Level Service Guide
6
CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)
The Agilent 53131A Calibration Procedures
To Calibrate the Offset for Channels 1 and 2
1 To access the Utility menu, power down, and press and hold the Recall (Utility)
key, then press POWER key.
2 Press Recall (Utility) key until CAL: OFFS 1 ? is displayed.
3 Press Enter key.
The display will ask you to remove any signals connected to Channel 1. Follow the
instruction.
4 Press Enter key.
The Counter momentarily displays CALIBRATING, and then it should display
OFFS 1 PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
5 To calibrate Channel 2, press any one of the arrow keys until CAL: OFFS 2 ? is
displayed.
6 Press Enter key.
The display will ask you to remove any signals connected to Channel 2. Follow the
instruction.
7 Press Enter key.
The Counter momentarily displays CALIBRATING, and then it should display
OFFS 2 PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
This completes the Offset Calibration procedure. Proceed to the next procedure to
calibrate the gain.
Zobrazit stránku 211
1 2 ... 207 208 209 210 211 212 213 214 215 216 217 ... 251 252

Komentáře k této Příručce

Žádné komentáře