DLYSWP Delay Sweep
DLYSWP Delay Sweep
Syntax
Figure 7-97 DLYSWP Syntax
Description
DLYSWP delays the start of the sweep until the specified time elapses
after the trigger event. With Option 007, and when using sweep times
<30 ms, the delay function can make the sweep start before the trigger
event. Executing "DLYSWP n;", where n is a non-zero number, is the
same as executing "DLYSWP ON;". Executing "DLYSWP 0;" is the same
as executing "DLYSWP OFF;". A positive DLYSWP value causes a
delayed sweep and a negative DLYSWP value causes the sweep to
pre-trigger. Querying DLYSWP either returns the value of sweep delay
length in seconds, or a "0" indicating the delay sweep is turned off.