Chapter 2 57
Making Measurements
Example 2: Improving Amplitude Measurements with Ampcor
Figure 2-6 Ampcor Measurement Setup
Set up the measurement.
1. Zero and calibrate the power meter and power sensor.
2. Connect the source output to the power splitter input. Connect the
system cable from the spectrum analyzer input to one of the power
splitter outputs. Connect the power sensor to the other power
splitter output. See Figure 2-6.
3. Set the source output to:
CW ............................................................................1.0 GHz
Amplitude................................................................... 0 dBm
4. On the power meter, enter the 1 GHz power sensor calibration factor.
5. The correction values should be determined using the same
measurement setup as the intended measurement will use. For the
purpose of this example, we will use the following spectrum analyzer
setup:
Center frequency......................................................1.0 GHz
Frequency span .........................................................5 MHz
Resolution bandwidth ............................................ 300 kHz
Log scale .......................................................... 1 dB/division
Zobrazit stránku 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 715 716

Komentáře k této Příručce

Žádné komentáře

Case-mate Non manuels d'utilisation

La liste de Case-mate Non produits contient 1 manuels d'utilisation pour les modèles 1. Tous les guides d'utilisation en format pdf peuvent être téléchargés ou consultés en ligne.

Cliquez sur un alphabet ci-dessous pour faire défiler la liste complète des modèles commençant par cette lettre

Modèles Type de document
CM010092
Fiche technique   Case-mate iPhone 3G/3GS Battery Case, 6 pages