Chapter 6: Replacing Assemblies
To remove and replace the front panel assembly
108
To remove and replace the front panel assembly
Use this procedure to remove and replace the front panel assembly. When necessary, refer to
other removal procedures. The graphics in this chapter are representative of the oscilloscope at
the time of this printing. Your unit may look different.
1
Disconnect the power cable and remove top and bottom covers.
2 Remove the AutoProbe assembly and Mylar flex cable (see page 104).
3 Remove the internal digital cable if you have an MSO model oscilloscope (see
page 106).
4 Use a 9/16 inch nut-driver to remove the hex nuts that secure the BNC connectors to
the front panel.
Figure 6-13
Removing the BNC nuts