Agilent Technologies 8000 Series Specifikace Strana 134

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 162
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 133
Chapter 6: Replacing Assemblies
To remove and replace the fan controller board
134
To remove and replace the fan controller board
Use this procedure to remove and replace the fan controller board. When necessary, refer to
other removal procedures. The graphics in this chapter are representative of the oscilloscope at
the time of this printing. Your unit may look different.
1
Disconnect the power cable and remove the bottom cover.
2 Disconnect the following cables from the fan control board.
Two fan power cables
Sense cable
Power harness cable
Control cable
Motherboard switch cable
3
Remove the two Torx T10 screws from the fan controller board.
4 Lift board out of chassis.
5 To re-install, reverse this procedure.
Figure 6-34
Removing the fan controller board
Control
cable
Power
harness
cable
Sense cable
Fan power
cables
Fan control
board
T10 screws (2)
Zobrazit stránku 133
1 2 ... 129 130 131 132 133 134 135 136 137 138 139 ... 161 162

Komentáře k této Příručce