Chapter 3: Testing Performance
To test delta time measurement accuracy
61
To test delta time measurement accuracy
Description
DTMA (delta time measurement accuracy) refers to the accuracy of time delay measurements
made between two edges of a sine wave acquired on a single scope channel. DTMA depends
heavily on whether the waveform is averaged, and is expressed as both an absolute error (peak)
and a standard deviation of multiple measurements (rms).
Specifications
The DTMA specification is shown below. The periods measured by the oscilloscope must fall
within the expected value of the period ± the value shown in the table. Limits are shown in the
Performance Test Record beginning on page 64.