14
Program and Interface Capabilities cont’d
Ethernet: 10/100Base-T
USB2.0: USB-TMC488 protocol (Rear x 1)
GPIB: IEEE-488.2 compliant
USB file system
USB 2.0 high-speed mass storage (MSC) class device (Front x 1)
Digital I/O interface
Connector: 25-pin female D.
Input/output pins: 14 open drain I/O bits
Absolute max input voltage: 5.25 V
Absolute min input voltage: - 0.25 V
Max logic low input voltage: 0.8 V
Min logic high input voltage: 2.0 V
Max source current: 1 mA, Vout = 0 V
Max sink current: 50 mA, Vout = 5 V
5 V power supply pin:
Limited to 600 mA, solid state fuse protected.
Safety interlock pin:
One active high pin and one active low pin. Activation of both pin enables
output voltage > 42 V.
Maximum Number of Simultaneously Triggered Units (using Digital I/O)
1
: 8
Furnished accessories
Power cable
USB cable
Quick Reference (English)
Product Reference CD-ROM (including PDF manuals, Agilent B2900A Quick I/V
Measurement Software and drivers)
Agilent I/O Library Suite
Software and drivers
Agilent B2900A Quick I/V Measurement Software
The Agilent B2900A series includes PC-based Agilent B2900A Quick I/V
Measurement Software. This powerful software makes it easy to quickly setup
and perform I/V measurements and to display the measurement data in a table
or graph without the need to perform any programming. The software allows
you to control up to four SMU channels over a GPIB or LAN connection, or one
B2900A series unit via a USB connection.
Operating environment:
Operating System:
Windows 7 (64bit/32bit), XP SP3 (32-bit).
Other requirements:
Microsoft .NET framework 4.0 or greater, and IO Libraries 16.0 or greater.