Slew rate/small signal bandwidth
Voltage Range Load
Resistance
Current range 200 mV 2 V 20 V 200 V
1 mA 44 mV/µs, 28 kHz 57 mV/µs, 18 kHz 57 mV/µs, 28 kHz 57 mV/µs, 28 kHz 10 MΩ
10 mA 44 mV/µs, 28 kHz 360 mV/µs, 20 kHz 360 mV/µs, 17 kHz 360 mV/µs, 28 kHz 10 MΩ
100 mA 28 mV/µs, 28 kHz 28 mV/µs, 20 kHz 28 mV/µs, 28 kHz 57 mV/µs, 28 kHz 10 MΩ
1 A 25 mV/µs, 28 kHz 25 mV/µs, 28 kHz 25 mV/µs, 28 kHz 10 MΩ
1.5 A 36 mV/µs, 28 kHz 36 mV/µs, 18 kHz 36 mV/µs, 28 kHz 10 MΩ
3 A 27 mV/µs, 28 kHz 27 mV/µs, 28 kHz 27 mV/µs, 28 kHz 10 MΩ
Noise 10 Hz to 20 MHz : < 3 mVrms, 20 V range, without external filter
V source overshoot:
< ±(0.1 %+10 mV). Step is 10 % to 90 % range, resistive load.
Voltage source range change overshoot:
≤ 250 mV. 100 kΩ load, 20 MHz bandwidth
Line regulation/load regulation: Included in voltage source specifications.
Load transient recovery time
Time to recover to within the settling band following a load change.
100 mA current change
1
800 mA current change
2
Settling band ±20 mV ±20 mV
Time 10 µs 30 µs
Current Range Settling Time
10 nA <10 ms
100 nA <10 ms
1 µA <500 µs
10 µA <250 µs
100 µA <250 µs
1 mA <80 µs
10 mA <80 µs
100 mA <80 µs
1 A <80 µs
1.5 A <80 µs
3 A <80 µs
1. At a load change to change the flowing current from +50 mA to +150 mA, 10 V
voltage force, 20 V range, 1 A limit.
2. At a load change to change the flowing current from +100 mA to +900 mA, 10 V
voltage force, 20 V range, 1.5 A limit.
Current Source:
Settling time:
Time required to reach within 0.1 % (0.3 % for 3 A range) of final value at
short condition.
Step is 10 % to 90 % range.
Note: Slew rate and small signal bandwidth can be down to -20% in maximum.