Addressing
Set the instrument address by using the front panel controls on the
oscilloscope after the GPIB interface has been installed on the rear panel of
the oscilloscope.
1 Press
Print/Utility , then press the I/O Menu softkey.
2 Press the Inst Addr softkey to select the instrument address. Increment
the address by successively pressing the Inst Addr softkey. The address
can also be incremented or decremented by turning the knob closest to
the
Cursors key.
Each device on the GPIB resides at a particular address, ranging from 0 to
30.
The active controller specifies which devices talk and which listen.
An instrument may be talk addressed, listen addressed, or unaddressed by
the controller.
If the controller addresses the instrument to talk, the instrument remains
configured to talk until it receives an interface clear message (IFC), another
instrument’s talk address (OTA), its own listen address (MLA), or a universal
untalk command (UNT).
If the controller addresses the instrument to listen, the instrument remains
configured to listen until it receives an interface clear message (IFC), its own
talk address (MTA), or a universal unlisten command (UNL).
Programming over GPIB
Addressing
3–4