Command Query Options and Query Returns
n/a :EXTernal:SETup? For the 54610:
EXTernal:OFFSet <offset_value>; COUPling {DC | AC | GND};
PROBe {X1 | X10 | X20 | X100}; PMODe {AUTo | MANual};
INPut {FIFTy | ONEMeg}; PROTect {OFF | ON}; SKEW <skew_value>
For the 54615/16:
EXTernal:COUPling {DC | AC | GND}; PROBe {X1 | X10 | X20 | X100};
PMODe {AUTo | MANual}; INPut {FIFTy | ONEMeg}; PROTect {OFF | ON}
:EXTernal:SKEW
<skew_value>
:EXTernal:SKEW? <skew_value> ::= external trigger skew value in seconds in NR3 format
:FUNCtion2:CENTer
<frequency>
:FUNCtion2:CENTer? <frequency> ::= the current center frequency in NR3 format.
The range of legal values is from 0 Hz to 10.00 GHz.
:FUNCtion2:MOVE
{LEFT)
n/a n/a
:FUNCtion<N>:OFFSet
<offset>
:FUNCtion<N>:OFFSet? <offset> ::= the value at center screen in NR3 format.
The range of legal values is +-10 times the current sensitivity of the
selected function.
<N> ::= 1 or 2
:FUNCtion<N>:OPERation
<operation>
:FUNCtion<N>:OPERation? <operation> ::=
{ADD | SUBTract | MULTiply} for :FUNCtion1:OPERation
{INTegrate | DIFFerentiate | FFT} for :FUNCtion2:OPERation
<N> ::= 1 or 2
n/a :FUNCtion2:PEAKs?
{FREQ1 | DB1 | FREQ2 |
DB2}
<measurement> ::= {FREQ1 | DB1 | FREQ2 | DB2}.
The measurement is the value of the peak specified in NR3 format.
:FUNCtion<N>:RANGe
<range>
:FUNCtion<N>:RANGe? <range> ::= the full-scale vertical axis value in NR3 format.
The range for FUNCtion1 is 8E-6 to 8E+6.
The range for the INTegrate function is 8E-9 to 400E+3.
The range for the DIFFerentiate function is 8E-6 to 1.6E11.
The range for the FFT function is 8 to 400 dB/div.
<N> ::= 1 or 2
:FUNCtion2:REFerence
<level>
:FUNCtion2:REFerence? <level> ::= the current reference level in NR3 format.
The range of legal values is from -160.0 dBV to +240.0 dBV in
increments of 2.5 dBV.
:FUNCtion2:SOURce
{CHANnel1 | CHANnel2 |
FUNCtion1}
:FUNCtion2:SOURce? {CHANnel1 | CHANnel2 | FUNCtion1}.
The current reference level value is in NR3 format. The range of
legal values is from -160.0 dBV to +240.0 dBV in increments of 2.5 dBV.
:FUNCtion2:SPAN
<span>
:FUNCtion2:SPAN? <span> ::= the current frequency span in NR3 format.
Legal values are 1.221 Hz to 9.766 Ghz
:FUNCtion<N>:VIEW
{ON | OFF}
:FUNCtion<N>:VIEW? {ON | OFF}
<N> ::= 1 or 2
:FUNCtion2:WINDow
{RECTangular | HANNing |
FLATtop | EXPonent}
n/a {RECTangular | HANNing | FLATtop | EXPonent}