Command Query Options and Query Returns
:MEASure:FALLtime :MEASure:FALLtime? <return_value> ::= time in seconds between the 10% and 90% voltage
levels in NR3 format
:MEASure:FREQuency :MEASure:FREQuency? <return_value> ::= frequency in Hertz in NR3 format
:MEASure:LOWer
<voltage>
:MEASure:LOWer? <voltage> ::= the user-defined lower threshold in volts in NR3 format
:MEASure:NWIDth :MEASure:NWIDth? <return_value> ::= negative pulse width in seconds in NR3 format
:MEASure:OVERshoot :MEASure:OVERshoot? <voltage> ::= the percent of the overshoot of the selected waveform in
NR3 format
:MEASure:PERiod :MEASure:PERiod? <return_value> ::= waveform period in seconds in NR3 format
:MEASure:PHASe :MEASure:PHASe? <return_value> ::= the phase angle value in degrees in NR3 format
:MEASure:PREShoot :MEASure:PREShoot? <return_value> ::= the percent of preshoot of the selected waveform in
NR3 format
:MEASure:PSTArt
<value>
:MEASure:PSTArt? <value> ::= the relative position of time marker 1 in degrees in NR3
format
:MEASure:PSTOp
<value>
:MEASure:PSTOp? <value> ::= the relative position of time marker 2 in degrees in NR3
format
:MEASure:PWIDth :MEASure:PWIDth? <return_value> ::= width of positive pulse in seconds in NR3 format
:MEASure:RISEtime :MEASure: RISEtime? <return_value> ::= rise time in seconds in NR3 format
:MEASure:SCRatch n/a n/a
:MEASure:SET100 n/a n/a
:MEASure:SET360 n/a n/a
:MEASure:SHOW
{ON | OFF}
:MEASure:SHOW? {ON | OFF}
:MEASure:SOURce
CHANnel <n>
:MEASure:SOURce? <n> ::=
1 or 2; an integer in NR1 format for 54600/03/10/15/16
1, 2, 3 or 4; an integer in NR1 format for the 54601/02
n/a :MEASure:TDELta? <return_value> ::= time difference in seconds between start and stop
markers in NR3 format
:MEASure:THResholds
{T1090 | T2080 | VOLTage}
:MEASure:THResholds? {T1090 | T2080 | VOLTage}
:MEASure:TSTArt
<value>
:MEASure:TSTArt? <value> ::= time at the start marker in seconds in NR3 format
:MEASure:TSTOp
<value>
:MEASure:TSTOp? <value> ::= time at the stop marker in seconds in NR3 format
n/a :MEASure:TVOLt
<tvolt_argument>,
<slope><occurrence>
<tvolt_argument> ::= positive or negative voltage level that the
waveform
must cross.
<slope> ::= direction of the waveform when <tvolt_argument> is
crossed.
<occurrence> ::= number of crossings to be reported.
<return_value> ::= time in seconds of specified voltage crossing in
NR3 format