Agilent OL Instrument Controller Installation and Configuration Guide
this button. A dialog box appears asking you if you want to close the window. Click Yes. 9. Immediately disconnect the crossover cable from the Agile
Figure 2. AIC 5000 Instrument and SS420x Connections Plug the SS420x power supply into an appropriate electrical outlet. Connect the power cord from
If you are using an instrument with an analog signal, you need to connect it to an SS420x attached to the Agilent Instrument Controller using the conn
1. From OL, select the Administration tab and then click Global Administration, followed by Agilent Instrument Controller. 2. Double-click the name
Baud Rate The communication baud rate is displayed. To change the baud rate, click Set Baud Rate. Base Frequency Select the base frequency. The defau
Configure Instruments Before you can use the Agilent Instrument Controller to acquire data, each instrument must be configured using the Agilent OL s
Important notes for the Reset to Factory Adapter • The Reset to Factory Adapter will cause a reset to factory defaults on any AIC, not just the one i
6 AIC 5000 Specifications Dimensions Height 3.95 inches 10.3 cm Width 13.3 inches 33.78 cm Depth 15.1 inches 38.35 cm Approximate Weight
Note This system utilizes a passive power factor corrected power supply when used in the 230V mode. This allows the system to pass the CE mark requi
7 Installing AIC Revision B for Agilent OL Use this procedure for the Agilent Instrument Controller Revision B. Revision B of the AIC is silver in
Notices Copyright © Scientific Software, Inc 1997-2005 © Agilent Technologies, Inc. 2005-2007. No part of this manual may be reproduced in any form or
Figure 1. AIC Connectors for Revision B 7. Power on the Agilent Instrument Controller and wait 5 minutes for initialization. The AIC will “beep” wh
2. Log in as Admin with no password. (Admin is case-sensitive.) 3. On the General tab, enter the following information: Name: for addressing the A
several possibilities, given below. Once you have resolved the problem, go back to the beginning of this section and repeat the steps to set up the A
Note If you are using a SCSI instrument, it must be powered ON prior to connecting it to the Agilent Instrument Controller. If you are using an instr
3. Select each hardware component for this Agilent Instrument Controller by clicking the box adjacent to the component name. You may select as many
Base Frequency Select the base frequency. The default setting for the SS420x is 10 Hz, which gives optimal signal to noise, and can be used in count
Configure Instruments Before you can use the Agilent Instrument Controller to acquire data, each instrument must be configured using the OL software.
8 Installing AIC Revision A for Agilent OL Use this procedure to install Revision A of the Agilent Instrument Controller. Revision A is black in co
Figure 1. 1a Not used 1b USB connection 2a Not used 2b USB connection 3,4a Monitor connection 4b Serial port connection 5 Not used 6 Not used 9 N
Comment: any additional comment you want to make OL server URL: for example http://aspen. 4. On the Hardware tab, click the Configure link and make
Contents 1 Using This Guide... 1 Introduction ...
• The laboratory network cable was not connected soon enough after clicking Initialize (see steps 5-6, above). Install the SS420x If your instrument
details on how to connect an instrument to the SS420x, see the documentation provided with the SS420x. Finish Configuring the Instrument Controller 1
to confirm the installation. If this is not the first installation, go to the Advanced tab and click Restart. The system will display the “Installin
countries). If selected incorrectly, you may see line frequency noise on your chromatogram. For best performance, select a base frequency close to
30 Installing AIC Revision A
9 Using the Reset to Factory Adapter for AIC Rev A/B The Agilent Instrument Controller (AIC) (Rev A and B only) is shipped with a USB flash drive th
• If you remove the special file on the Reset to Factory Adapter or rename it, then the device will no longer cause a reset to factory. Don’t use th
10 Updating an AIC This section applies only to the Revision A and Revision B versions of the Agilent Instrument Controller. For support purposes, i
34 Updating an AIC
12 Specifications for AIC Rev A and B Agilent Instrument Controller Revisions A and B Environment Operating temperature: 0oC to 40oC (32oF to 10
Install the SS420x...18 Connect the Instruments ...
Printed in the U.S.A. February, 2007 *G4650-90017* G4650-90017
1 Using This Guide Introduction This guide describes how to install the Agilent Instrument Controller (AIC) for use with Agilent OL. The Agilent In
Note Place the AIC 5000 on a flat, stable surface. You can position the unit in the upright position using the included stand, or in the horizontal p
3 Installing AIC 5000 for Agilent OL Use this procedure for installing and configuring the Agilent Instrument Controller AIC 5000. The AIC 5000 is b
5. Disconnect the computer from the network and change the TCP/IP settings to use the static IP address 192.168.1.2. Contact a network administrator
Configure the Agilent Instrument Controller Device 1. From the configuration computer (see “Before You Start” section), open Internet Explorer and g