2- 30 Agilent B2900 Users Guide, Edition 2
Introduction
Operation and Functions
Source/Measurement Capabilities
Sweep Measurement
Number of steps: 1 to 2500
Sweep mode: linear, logarithmic (log), or list
Sweep direction: single or double
Type: DC or pulse
Minimum programmable value to create list sweep waveform:
B2901A/B2902A: 20 s with 1 s resolution
B2911A/B2912A: 10 s with 1 s resolution
Digitizing/Sampling Measurement
Digitizing/sampling measurement can be performed by appropriate setting of
measure speed and trigger interval.
Minimum trigger interval (maximum speed of measurement):
B2901A/B2902A: 20 s (50000 points/s)
B2911A/B2912A: 10 s (100000 points/s)
Data Buffers
Data buffers are array of measurement elements. Each elements hold the following
values.
Measured voltage
Measured current
Measured resistance
Time stamp
Measurement status
Source setting
Maximum buffer size: 100000 points/channel