Chapter 2 41
Performance Tests
E5061A Performance Test Record (for Option 150/250)
2. Performance Test
Direction: S21 (Phase)
Direction: S12 (Magnitude) (Option 250 only)
Direction: S12 (Phase) (Option 250 only)
Direction: S11 (Magnitude) (Option 150 only)
Direction: S11 (Phase) (Option 150 only)
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [deg
rms]
Test result [deg
rms]
300 k - 1 M 3 k 0.05
1 M - 100 M 3 k 0.03
100 M - 1.5 G 3 k 0.03
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [dB rms] Test result [dB rms]
300 k - 1 M 3 k 0.008
1 M - 100 M 3 k 0.005
100 M - 1.5 G 3 k 0.005
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [deg
rms]
Test result [deg
rms]
300 k - 1 M 3 k 0.05
1 M - 100 M 3 k 0.03
100 M - 1.5 G 3 k 0.03
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [dB rms] Test result [dB rms]
300 k - 1 M 3 k 0.008
1 M - 100 M 3 k 0.005
100 M - 1.5 G 3 k 0.005
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [deg
rms]
Test result [deg
rms]
300 k - 1 M 3 k 0.05
1 M - 100 M 3 k 0.03
100 M - 1.5 G 3 k 0.03