11. Limit Test
Step 4. Press Ripple Limit to display the Ripple Limit menu.
Step 5. Press
Ripple Test to turn the ripple test ON. You can display the limit lines on the screen
by turning on
Ripple Limit.
Initializing the limit table
The following operations initialize the limit table.
At power-on
When presetting
When calling a ripple table with zero frequency bands
•When
Clear Limit Table - OK is pressed in the Edit Ripple Limit menu
Softkey Function
Ripple Test Sets the ripple test ON/OFF.
Ripple Limit Sets the ripple limit line display ON/OFF.
Ripple Value Sets how the ripple values are displayed. Available settings are off,
absolute value (difference between maximum and minimum values
within the band) display, and margin (difference between absolute value
of ripple and ripple limit) display.
Ripple Value Band Selects the band for which you want to display the ripple value.
Edit Ripple Limit
Opens the ripple limit table for editing the ripple limit.
*1
*1.To use the ripple test function, you must first define the ripple limit. For more on how to
define the ripple limit, see “Configuring ripple limit” on page 387.