Agilent Technologies ENA Series Uživatelská příručka Strana 530

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 775
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 529
524 Chapter 15
Measurement Examples
Measuring Parameters with Cable
5. Setting the Time Domain Function
Set the conversion function to display the response in time the domain. If you enable this
setting, the response in time domain is displayed as shown in Figure 15-34. A peak
indicating a small mismatch appears at the location of the connector.
Figure 15-34 Response in time domain
Setting Description Key Operation
Data format: real
- Real
Setting the transformation type to low-pass
impulse
- Transform - Lowpass Imp.
Setting the window type to maximum.
Window - Maximum
Setting the display range: from 0 s to 10 ns
Start -
Stop -
Enabling the transformation function
Transform (set to ON)
Executing auto scale
- Auto Scale
Zobrazit stránku 529
1 2 ... 525 526 527 528 529 530 531 532 533 534 535 ... 774 775

Komentáře k této Příručce

Žádné komentáře