Replacing the 8753ES with the E5070B/E5071B
Important Functional Differences
F. Replacing the 8753ES
with the E5070B/E5071B
along with the minimum/maximum measurement value for each segment as retrieved items
in test results. On the other hand, the E5070B/E5071B supports pass/fail results,
measurement value, and upper/lower limit values for each measurement point as well as
pass/fail results of the active trace for each channel.
Analytical Functions
Although the 8753ES does not support the fixture simulator function, the E5070B/E5071B
does. The fixture simulator supported by the E5070B/E5071B include the
balanced-unbalanced transformation function for analyzing balanced devices, and the
matching circuit function.