Agilent Technologies ENA Series Uživatelská příručka Strana 521

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 775
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 520
Chapter 15 515
Measurement Examples
Measuring an Unbalanced and Balanced Bandpass Filter
15. Measurement Examples
2. Setting the Measurement Conditions
Follow the procedure below to set the measurement conditions. The measurement
parameters for balanced measurements should be set after unbalanced-balanced
conversion. Here, set the measurement parameters for observing the characteristics
achieved during unbalanced measurements.
Figure 15-27 Results of unbalanced measurements
Setting Description Key Operation
Preset for setting
-
OK
Center frequency: 940 MHz
Frequency span: 200 MHz
Number of traces: 2
-
Num of Traces - 2
Trace-1 measurement parameter: S
21
- S21
Trace-2 measurement parameter: S
31
- S31
Allocate a trace to upper and lower displays
-
Allocate Traces -
Auto-scale all traces
-
Auto Scale All
Zobrazit stránku 520
1 2 ... 516 517 518 519 520 521 522 523 524 525 526 ... 774 775

Komentáře k této Příručce

Žádné komentáře