Agilent Technologies E5503A Uživatelská příručka Strana 153

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 465
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 152
7-28
Agilent Technologies E5500 Phase Noise Measurement System
Absolute Measurement Examples
Free-Running RF Oscillator
Selecting Loop
Suppression
Verification
1. From the Define menu, choose Measurement; then choose the Cal tab
from the Define Measurement window.
2. In the Cal dialog box, check Verify calculated phase locked loop
suppression and Always Show Suppression Graph. Select If limit is
exceeded: Show Loop Suppression Graph.
3. When you have completed these operations, click the Close button
.
Setup Considerations
for the Free-Running
RF Oscillator
Measurement
Measurement Noise Floor
The signal amplitude at the R input (Signal Input) port on the Agilent/HP
70420A sets the measurement noise floor level. Use the following graph to
determine the amplitude required to provide a noise floor level that is below
the expected noise floor of your UUT. (The Checking the Beatnote
procedure in this section will provide you with an opportunity to estimate
the measurement noise floor that your UUT will provide.)
Zobrazit stránku 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 464 465

Komentáře k této Příručce

Žádné komentáře