Agilent Technologies E5503A Uživatelská příručka Strana 278

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 465
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 277
9-10
Agilent Technologies E5500 Phase Noise Measurement System
Residual Measurement Examples
Amplifier Measurement Example
Residual Connect Diagram Example
Making the
Measurement
Calibrate the Measurement using Measured +/- DC Peak Voltage
Refer to Chapter 8, “Residual Measurement Fundamentals” for more
information about residual phase noise measurements calibration types.
Procedure
1. Connect circuit as per Figure 9-2 on page 9-11, and tighten all
connections.
2. Measure the power level that will be applied to the Signal Input port of
the Agilent/HP 70420As phase detector. The following chart shows the
acceptable amplitude ranges for the Agilent/HP 70420A phase detectors.
Zobrazit stránku 277
1 2 ... 273 274 275 276 277 278 279 280 281 282 283 ... 464 465

Komentáře k této Příručce

Žádné komentáře