Agilent Technologies E5503A Uživatelská příručka Strana 396

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 465
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 395
18-2
Agilent Technologies E5500 Phase Noise Measurement System
Reference Graphs and Tables
Approximate System Phase Noise Floor vs. R Port Signal Level
Approximate System Phase Noise Floor vs. R
Port Signal Level
The sensitivity of the phase noise measurement system can be improved by
increasing the signal power at the R input port (Signal Input) of the phase
detector in the test set. The graph shown above illustrates the approximate
noise floor of the Agilent/HP 70420A test set for a range of R input port
signal levels from -15 dBm to +15 dBm. These estimates of sensitivity
assume the signal level at the L port is appropriate for either the microwave
or the RF mixer that is used (+7 dBm or +15 dBm, respectively). The
approximate phase Detector calibration Constant that results from the input
signal level at the R port is shown on the right side of the graph.
Zobrazit stránku 395
1 2 ... 391 392 393 394 395 396 397 398 399 400 401 ... 464 465

Komentáře k této Příručce

Žádné komentáře