Agilent Technologies E5500 Phase Noise Measurement System
A-5
Connector Care and Preventive Maintenance
Using, Inspecting, and Cleaning RF Connectors
CAUTION Do not allow excessive alcohol to run into the connector. Excessive alcohol
entering the connector collects in pockets in the connectors internal parts.
The liquid will cause random changes in the connectors electrical
performance. If excessive alcohol gets into a connector, lay it aside to allow
the alcohol to evaporate. This may take up to three days. If you attach that
connector to another device it can take much longer for trapped alcohol to
evaporate.
Lint-Free cloths: 9310-4242
Small foam swabs: 9300-1270
Large foam swabs 9300-0468
Table A-2 Cleaning Supplies Available from Agilent Technologies
Product Part Number