Command Reference
Status Subsystem (:STATus)
Chapter 4194
:QUEStionable:CALibration:NTRansition
Supported PSG-A Series
:STATus:QUEStionable:CALibration:NTRansition <value>
:STATus:QUEStionable:CALibration:NTRansition?
This command determines what bits in the Data Questionable Calibration Condition
Register will set the corresponding bit in the Data Questionable Calibration Event
Register when that bit has a negative transition (1 to 0).
The variable <value> is the sum of the decimal values of the bits that you want to
enable.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to “Data Questionable Calibration Status Group” on page 133 for
more information.
:QUEStionable:CALibration:PTRansition
Supported PSG-A Series
:STATus:QUEStionable:CALibration:PTRansition <value>
:STATus:QUEStionable:CALibration:PTRansition?
This command determines what bits in the Data Questionable Calibration Condition
Register will set the corresponding bit in the Data Questionable Calibration Event
Register when that bit has a positive transition (0 to 1).
The variable <value> is the sum of the decimal values of the bits that you want to
enable.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to “Data Questionable Calibration Status Group” on page 133 for
more information.