To transfer an individual mask template
The HP 54653A ScopeLink software 2.0 can transfer the setup of an
individual mask template from the oscilloscope to the PC and store it as a file.
The file can be used as a backup of the individual mask template. To restore
the individual mask template, simply send the file back to the oscilloscope.
The file can also be used as a part of a test library. A test sequence can be
constructed by using individual steps and mask templates stored in the PC.
1 Select Data Transfer from the menu bar, then select Test
Automation.
2 When the Test Automation Panel appears, check the box labelled
Mask for transferring an individual mask template.
3 Select the box trailing the label number, and enter the mask template
number that you want to work on.
Valid mask template numbers are from 1 to 40.
4 Enter the filename into the Filename box.
The extension of the filename must be ".SLM", which is automatically
appended to the filename if no extension is given.
5 To retrieve an individual mask template, check the box labelled
Receive.
To restore an individual mask template, check the box labelled
Send.
6 Select Okay to start the transfer. To return to the main menu, select
Quit.