To create a label for a front-panel setup
Each front-panel setup stored in Test Automation Module can have a label up
to 60 characters long to identify the setup.
1 Press Print/Utility .
2 Press the Sequence Admin softkey.
The Test Automation Module remembers which menu you were in last.
If you cannot find the Sequence Admin softkey,
press Print/Utility again to return to the root level of the menu
tree. Then press the Sequence Admin softkey.
3 Press the Define Sequence softkey.
4 Press the Step softkey, or turn the knob closest to the Cursors key,
to select the step number for adding the user message.
5 Press the Define Step softkey.
6 Press the Messages softkey.
7 Press the Label softkey.
A softkey menu with six softkey choices appears. Five of them are related to
editing the message.
•
Enter Enters the highlighted character in the character area into the
message area under the cursor position.
•
Space Enters a space into the message area under the cursor position.
•
Delete Deletes the character in the message area under the cursor
position.
•
Insert Toggles the insert function on or off. With insert on, pressing
Enter or Space inserts the character or space into the message before the
character at the cursor position. The rest of the message is shifted to the
right by one character. With insert off, pressing Enter or Space replaces
the character highlighted in the character area.
•
Clear To End Clears the characters from the cursor position to the end
of the message including the character under the cursor.
•
Done Exits the editing of the message.
Operating the Test Automation Module
To create a label for a front-panel setup
24