To edit an individual pixel of a mask
1 Press Print/Utility twice.
2 Press the Sequence Admin softkey.
3 Press the Mask Editor softkey.
4 To select the mask to be modified, press the Edit Mask softkey, or
turn the channel 1 position knob.
5 Toggle the Edit Line softkey to select the limit line that you want to
edit.
6 Turn the Delay knob to move the cursor to the pixel that you want to
modify.
7 Turn the knob closest to the Cursors key to edit the vertical position
of the pixel.
Pixel editing operating hint
The time and voltage shown at the bottom of the screen corresponds to the
current time base and vertical setting of lowest number of all active channels.
If the mask is voltage and time dependent, make sure that the current time base
and vertical setting are the same as the one that you are going to use during
the actual testing.
Operating the Test Automation Module
To edit an individual pixel of a mask
33