Agilent Technologies HP 8642A/B Specifikace Strana 125

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 453
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 124
Considerations for the Loop Suppression Verification
Verification of the loop suppression may be prevented by excessive noise
close to the carrier, by a large spur within the bandwidth of the PLL, or
by injection locking.
It should be assumed that an indicated Accuracy Specification Degradation
applies at all offset frequencies (both inside and outside of the loop
bandwidth).
If the Detector Constant and VCO Tune Constant are known to be
accurate, the PTR is less than 10 MHz, and the VCO tuning port
bandwidth is less than PLL BW x 10, then a measurement without Loop
Suppression Verification will be accurate.
2-56 Calibr Process: 3>N w/PLL
Zobrazit stránku 124
1 2 ... 120 121 122 123 124 125 126 127 128 129 130 ... 452 453

Komentáře k této Příručce

Žádné komentáře