Table 8-5. Status Messages (8 of 11)
Returning to 'Main Software Level'... The System gave-up and re-booted itself.
Action:
Verify that the measurement parameters are correctly defined for the measurement.
RF SPECTRUM ANALYZER MODEL HP3585A REQUIRED... An HP 3585A/B is required for
calibration of the HP 11848A from 100 kHz to 40 MHz.
Action:
Verify that the HP 3585A/B is correctly connected to the HP-IB interface, and that it is
listed in the System Configuration Table. (Other RF analyzers supported by the HP 3048A can be
used to update the calibration files if an HP 3585A is not available, Refer to RF Analyzers
in Chapter 6, System Configuration for a list of supported RF Analyzers.)
Selecting the LNA... The System is determining if the Low Noise Amplifier is required.
Setting up for A4C32 adjustment... The System is processing part of the
internal adjustment.
Setting up HP 11848A for A4R51 Adjustment... The System is processing part of the Internal
Adjustment.
Setting up measurement configuration... The System is configuring the instruments and/or
the HP 11848A internal paths for the measurement.
Solving for (Osc unknown)... The System is computing the noise of a source.
Solving for Osc_A,B»or C.The system is computing the noise of each of the sources.
Sorting config table... The System is alphabetizing the entries in the System Configuration
Table.
Span too narrow. Cannot integrate... The specified Range of Integration is not great
enough to span a bin of the noise trace.
Action:
Increase the specified Range of Integration.
Storing HP 11848A calibration data to disc at ()... This message indicates that a disc
write operation is in progress for the Calibration Data.
Storing System Configuration... This message indicates that a disc write operation is in
progress for the system configuration.
8-36 Status Messages