Chapter 6: Replacing Assemblies
To remove and replace the front panel bezzle, front panel keyboard, touchscreen, and Autoprobe
board
126
Figure 6-20
Figure 6-21
16
Disconnect the touchscreen cable (the one with the blue connector) and the three display cables
(the ones with the white connectors) as shown in Figure 6-20 and Figure 6-21.
The touchscreen is
mounted on the other
side of this sheet metal
Main oscilloscope
body
Disconnect this cable (the one with the blue connector) because it
is connected to the touchscreen
Remove these two
display cables
(white connectors)
from the inverter
board because they
are connected to the
touchscreen display
Also disconnect this
display cable
These are the three
cables mentioned in
the figure immediately
above that need
disconnected before
you remove the
touchscreen display