Setting Up the Oscilloscope
To connect oscilloscope probes
2-31
2 Connect the probe to the circuit of interest using the browser or other probing
accessories.
Figure 0-4
Probing the Circuit
3 To disconnect the probe, push the small latch on top of the probe connector
to the left, then pull the connector body away from the front panel of the
oscilloscope without twisting it.
CAUTION Do not attempt to twist the snap-on probes on or off the oscilloscope’s BNC connector.
Twisting the probe connector body will damage it.
CAUTION For the 9000A series oscilloscopes do not exceed the maximum input voltage
rating. The maximum input voltage for the 50 inputs is ±5 Vpeak, CAT I and for
the 1 M inputs, it is 150 Vrms (or dc), CAT I, ±250 V (dc + ac) in ac coupling.
!