326 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:XUNits:USE
(see page 790)
Command Syntax
:MARKer:XUNits:USE
When DEGRees is selected for :MARKer:XUNits, the :MARKer:XUNits:USE
command sets the current X1 location as 0 degrees and the current X2
location as 360 degrees.
When PERCent is selected for :MARKer:XUNits, the :MARKer:XUNits:USE
command sets the current X1 location as 0 percent and the current X2
location as 100 percent.
Once the 0 and 360 degree or 0 and 100 percent locations are set, inputs
to and outputs from the :MARKer:X1Position, :MARKer:X2Position, and
:MARKer:XDELta commands/queries are relative to the set locations.
See Also "Introduction to :MARKer Commands" on page 318
":MARKer:XUNits" on page 325
":MARKer:X1Y1source" on page 321
":MARKer:X2Y2source" on page 323
":MEASure:SOURce" on page 361
":MARKer:X1Position" on page 320
":MARKer:X2Position" on page 322
":MARKer:XDELta" on page 324