:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 497
:SBUS<n>:SPI:CLOCk:TIMeout
(see page 790)
Command Syntax
:SBUS<n>:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR3 format
The :SBUS<n>:SPI:CLOCk:TIMeout command sets the SPI signal clock
timeout resource in seconds from 100 ns to 10 s when the
:SBUS<n>:SPI:FRAMing command is set to TIMeout. The timer is used to
frame a signal by a clock timeout.
Query Syntax
:SBUS<n>:SPI:CLOCk:TIMeout?
The :SBUS<n>:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also "Introduction to :TRIGger Commands" on page 589
":SBUS<n>:SPI:CLOCk:SLOPe" on page 496
":SBUS<n>:SPI:SOURce:CLOCk" on page 499
":SBUS<n>:SPI:FRAMing" on page 498