Obsolete and Discontinued Commands 33
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 713
:CHANnel<n>:PMODe
(see page 790)
Command Syntax
:CHANnel<n>:PMODe <pmode value>
<pmode value> ::= {AUTo | MANual}
<n> ::= 1 to (# analog channels) in NR1 format
The probe sense mode is controlled internally and cannot be set. If a
probe with sense is connected to the specified channel, auto sensing is
enabled; otherwise, the mode is manual.
If the PMODe sent matches the oscilloscope's setting, the command will be
accepted. Otherwise, a setting conflict error is generated.
Query Syntax
:CHANnel<n>:PMODe?
The :CHANnel<n>:PMODe? query returns AUT if an autosense probe is
attached and MAN otherwise.
Return Format
<pmode value><NL>
<pmode value> ::= {AUT | MAN}
NOTE
The :CHANnel<n>:PMODe command is an obsolete command provided for compatibility to
previous oscilloscopes.