Chapter 10 329
Specifications and Supplemental Performance Characteristics
Basic Characteristics
10. Specifications and
Supplemental Performance
Characteristics
Measurement Time
Figure 10-2 Measurement Time (SPC)
NOTE When the 42941A Impedance Probe or the 42942A Terminal Adapter is used,
measurement time is 1.5 times longer than the value shown in Figure 10-2.
Trigger Function
Measurement Bandwidth/Averaging
Trigger types Continuous, single, number of groups
Trigger sources Internal (free run), external (BNC connector input),
GP-IB or LAN, manual (front key)
Trigger event types Point trigger, sweep trigger
Measurement bandwidth
Range 1 (Fast) to 5 (Precise), 5 steps
Averaging
Types Sweep-to-sweep averaging, point averaging