334 Chapter 10
Specifications and Supplemental Performance Characteristics
Basic Characteristics
NOTE SPC at frequency > 10 MHz.
K
BW
=
at measurement BW = 5 1
at measurement BW = 4 1
at measurement BW = 3
frequency £ 1 MHz 3
frequency > 1 MHz 4
at measurement BW = 2
frequency £ 1 MHz 4
frequency > 1 MHz 5
at measurement BW = 1
frequency £ 1 MHz 6
frequency > 1 MHz 10
Y
ODC
=
at dc bias range = 1 mA 0 [S]
at dc bias range = 10 mA 1 [mS]
at dc bias range = 100 mA 10 [mS]
E
PL
[%] =
when 16048G is used
when 16048H is used
Y
OL
=
when 16048G is used
[nS]
when 16048H is used
[mS]
Z
SL
=
when 16048G or
16048H is used
frequency ³ 500 Hz 2 [mW]
frequency < 500 Hz 5 [mW]
0.02 2
f
100
---------
´+
0.02 3
f
100
---------
´+
500
f
100
---------
´
1
f
100
---------
´