Agilent Technologies 8935 Series Uživatelský manuál Strana 106

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 352
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 105
106 Chapter 1
GPIB Command Dictionary
CANalyzer subsystem
:PATH:GAIN <string>
:PATH:GAIN?
These commands set/query the gain of the CDMA analyzer. This
command will only be enabled when autoranging is turned off
(CAN:PATH:MODE ‘Hold’).
This command may conflict with the code domain analyzers settings
CDAN:PATH:GAIN and CDAN:PATH:MODE. For example, if
autoranging is turned off (‘Hold’) in the CDMA analyzer and turned on
in the Code Domain Analyzer (‘Auto’) the ‘Hold’ setting will always
override the ‘Auto’ setting. When autoranging is turned off, setting
either analyzers gain will automatically change the others to the same
setting.
Use the CAN:POW:GAIN command when making an average power
measurement.
Syntax
CAN:PATH:GAIN ‘0 dB’
‘6 dB’
‘12 dB’
‘18 dB’
‘24 dB’
‘30 dB’
‘36 dB’
CAN:PATH:GAIN?
Screen/field equivalent
CAN:PATH:GAIN sets the lower subfield of the Gain field of the CDMA
ANALYZER screen. This subfield is only selectable when the upper
subfield is set to Hold.
Zobrazit stránku 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 351 352

Komentáře k této Příručce

Žádné komentáře