Timing Behavior
Table 15 details the ways in which timing behavior is different.
Table 15 Timing Behavior Changes
Change Affects
Command execution may be different. Both
In all mainframes and modules firmware revisions before 3.x GPIB will
block during command execution, except when executing functions,
such as logging and sweep, that don’t tolerate blocking. This is
identical to the behavior of the 8167/8. A side effect of this is that
*OPC? always returns 1.
In later firmware revisions GPIB only blocks if a command can't be
processed because of a pending command. While a command is
pending *OPC? returns 0 now. This is the behaviour of the 8153.
Both
When continuous triggering and averaging times are greater than 1
second, the read-out values reset after the averaging time is over; there
is no sliding behavior.
8153