2-4 8510XF Network Analyzer Systems
Installation
Site Preparation
Weight and dimensions
The table below shows the maximum weight and dimensions of the 8510XF
system, as installed in the system rack, with the test heads on an attached
work surface.
Table 2-3 System Weight and Dimensions
The table below shows the weight and dimensions of a single test head (left
or right):
Figure 2-1 Testhead Footprint Dimensions for Mounting (left or right)
Weight Height Width Depth
280.3 kg (618 lbs) 162 cm (63.8 in) Rack only: 60 cm (23.6 in)
Plus work surface: 100 cm (39.4 in)
Rack Only: 90.5 cm (35.6 in)
Plus work surface: 145.4 cm (57.3 in)
Table 2-4 Test Head Weight and Dimensions
Test Head Model Weight Height Width Depth
E7342 L/R 9.07 kg
(20 lb)
18.5 cm
(7.3 in)
24.7 cm
(9.7 in)
31.4 cm
(12.4 in)
E7352 L/R 9.75 kg
(21.5 lb)
18.5 cm
(7.3 in)
24.7 cm
(9.7 in)
31.4 cm
(12.4 in)