3-26 8510XF Network Analyzer Systems
Operation
Standard Types
Full 2-Port
Transmission
Standards
These standards are used to measure transmission and match for full 2-port
calibrations. When you perform a calibration of that type using a 1.0 mm
calibration kit, you will see the following menu:
Figure 3-14 Full 2-Port Transmission Calibration Menu
& Equivalent GPIB Commands
Notes on the Standards {FWD. TRANS.
THRU}
Make a “thru” connection (that is, connect one port to the
other through a cable).
{FWD.MATCH
THRU}
Make a “thru” connection (that is, connect one port to the
other through a cable).
{REV.TRANS.
THRU}
Make a “thru” connection (that is, connect one port to the
other through a cable).
{REV.MATCH
THRU}
Make a “thru” connection (that is, connect one port to the
other through a cable).
NOTE It is necessary to make all four of these measurements.