Chapter 13 251
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:MARK:FUNC:MULT:TARG
Syntax :CALCulate{[1]|2|3|4}[:SELected]:MARKer:FUNCtion: MULTi: TARGet <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:MARKer:FUNCtion: MULTi: TARGet?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), sets the target value to be searched with the multi
target search.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:MARK1:FUNC:MULT:TARG -12.5"
20 OUTPUT 717;":CALC1:MARK1:FUNC:MULT:TARG?"
30 ENTER 717;A
Related commands :CALC{1-4}:MARK:FUNC:MULT:TTR on page 253
:CALC{1-4}:MARK:FUNC:MULT:TYPE on page 254
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key
[Marker Search] - Multi Target - Target Value
<numeric>
Description Target value for target search
Range -5E8 to 5E8
Preset value 0
Unit Varies depending on the data format as follows:
Logarithmic Magnitude: dB (decibel)
Phase, Expand Phase, Positive Phase: ° (degree)
Group Delay: s (second)
Others: No unit