Agilent Technologies E5071C Uživatelská příručka Strana 426

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 551
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 425
426 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:SWE:TIME
Syntax :SENSe{[1]|2|3|4}:SWEep:TIME[:DATA] <numeric>
:SENSe{[1]|2|3|4}:SWEep:TIME[:DATA]?
Description Sets the sweep time of channel 1 (:SENS1) to channel 4 (:SENS4).
When the auto setting of the sweep time is ON, even if you try to set the sweep time to any
value with this command, it automatically returns to the value defined by the
E5061A/E5062A. Before using this command, turns OFF the auto setting of the sweep
time (specify OFF with the :SENS{1-4}:SWE:TIME:AUTO command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:SWE:TIME 1.5"
20 OUTPUT 717;":SENS1:SWE:TIME?"
30 ENTER 717;A
Related commands :SENS{1-4}:SWE:TIME:AUTO on page 427
Equivalent key
[Sweep Setup] - Sweep Time
<numeric>
Description Sweep time
Range Varies depending on the measurement conditions.
Preset value Varies depending on the measurement conditions.
Unit s (second)
Zobrazit stránku 425
1 2 ... 421 422 423 424 425 426 427 428 429 430 431 ... 550 551

Komentáře k této Příručce

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm