Chapter 4 65
Performing a Calibration
Calibration
4. Performing a Calibration
As shown in Table 4-2, you need to set different items depending on the standard type.
You need to set the items identified by marks in the table above.
Defining a Standard Class Assignment
To select the standard to be applied to the OPEN measurement for each port, use the
following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:OPEN on page 378
To select the standard to be applied to the SHORT measurement for each port, use the
following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:SHOR on page 379
To select the standard to be applied to the LOAD measurement for each port, use the
following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:LOAD on page 377
To select the standard to be applied to the THRU measurement between each pair of ports,
use the following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:THRU on page 380
Table 4-2 Settings Specific to Each Standard Type
Standard Types C0 to C3 L0 to L3
Offset
Delay
Offset
Loss
Offset Z0
Arbitrary
Impedance
OPEN √√√
SHORT √√√√
LOAD √√√
THRU √√√
Arbitrary
Impedance
√√√