Chapter 13 265
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:MARK{1-10}:BWID:DATA?
Syntax :CALCulate{[1]|2|3|4}[:SELected]:MARKer{[1]|2|3|4|5|6|7|8|9|10}:BWIDth:DATA?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), reads out the bandwidth search result of marker 1
(:MARK1) to marker 9 (:MARK9) and reference marker (:MARK10).
If the bandwidth search is impossible, an error occurs and the command is ignored. In this
case, no query response is obtained. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3},{numeric 4}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:MARK1:BWID:DATA?"
20 ENTER 717;A,B,C,D
Related commands :CALC{1-4}:MARK:BWID on page 243
:CALC{1-4}:MARK{1-10}:BWID:THR on page 266
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key No equivalent key is available on the front panel.
Description
{numeric 1} The bandwidth.
{numeric 2} The center frequency.
{numeric 3} The Q value.
{numeric 4} The loss.