Troubleshooting 5
16800 Series Portable Logic Analyzers Service Guide 107
If the listing does not look similar to the figure, then
there is a possible problem with the cable or probe tip
assembly. Causes for cable test failures include:
• Open channel.
• Channel shorted to a neighboring channel.
• Channel shorted to either ground or a supply voltage.
Connect and configure the logic analyzer to test other pods
1 On the E5383A Flying Lead Probe set you have been
using, move Clk1 to the pulse generator Channel 2
OUTPUT (from OUTPUT
).
2 Disconnect the logic analyzer Pod 1 cable from the flying
lead probe set and connect the logic analyzer Pod 2 cable.
3 Connect the second E5383A Flying Lead Probe set to the
logic analyzer Pod 1 cable.
4 Connect the second flying lead probe set’s Clk1 to the 2 x
9 test fixture at the pulse generator Channel 2 OUTPUT
.
5 In the Analyzer Setup dialog’s Buses/Signals tab,
un- assign all channels of Pod 1.
6 Assign all channels of the pod to be tested. If the pod to
be tested has a clock, assign it too (place a check mark in
the box under the channel name).
7 Select the Sampling tab.
8 Click the Thresholds and Sampling Positions... button.
9 Move the starting position to - 2 ns.