Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 127
To remove and replace the touch screen controller board
1 Perform previous procedures:
• “To prepare the instrument for disassembly" on
page 116.
• “To remove and replace the cover" on page 118.
• “To remove and replace the front panel assembly" on
page 125.
2 Disconnect the touch screen cable from the touch screen
controller board.
3 Using a Torx T10 screwdriver, remove the 2 screws securing
the board to the front panel board.
4 Reverse this procedure to install the touch screen controller
board.
Touch
screen
controller
board
T-10
screws
(2)
Touch screen
cable
Komentáře k této Příručce